GE Healthcare · CTLightSpeed
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
A Loud noise is coming from the table, even when it is not moving. A loud "moaning" sound is coming from the table. The sound is fairly constant and does not seem to vary when the table it elevated or the cradle is moved in and out. The sound seems to come from the base of the table, farthest away from the gantry.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2294322-3 (replace 2294322-2) : Table Elevation Linear Actuator, Zeus thru H16 Table. See long description for details. Part #: 2294322-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actiator, Linear, Zeus thru H16 Table, Pinned, R4 Ball Screw Fillet FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR2079949, 212008898, 5/19/09. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 2294322-3 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2079949 Part# 46-296561P1 converts to Part #: 2294322-2 in FEMC Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actuator, Linear, Zeus thru H16 Table, Pinned FRU: Yes Status: Orderable Parts Offered as: Current: 2294322-2 Harvested: Not Available Exchange: Not Available Comments: Added per ECR ms13452 041210
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required