Service Note 43788_1 Feb/08 Millennium MG Detector Brackest Inspection

Symptom

Service Note 43788_1 Feb/08 Millennium MG Detector Brackest Inspection Service Note 43788 _1 Feb/08 Millennium MG Detector Brackest Inspection Problem In an isolated case, it was found that four detector bracket bolts were loose. This caused the following problems: Difficulties in Collimator Exchange procedure Failure of the COR Callibration procedure Further evaluation revealed that the following problems could also be expected: SPECT Images – Jumps between the images of Detector 1 and Detector 2 during data processing. Knocking Noise – Knocking noise during rotation and/or radial movement.

Applies to

Models
Millennium
Product
MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine