GE Healthcare · MRSigna
Modality Work list Schedule window will not open Due to WL Files.
Symptom
Modality Work list Schedule window will not open following a power issue at the site Site experienced a power issue where the /usr/g/hisris/WLFilterFile and/or the /w/config/WLSystem.cfg files were corrupted. The issues with these files were corrected remotely, but the techs could not confirm the fix because they were scanning at the time. Ultimately, when they were able to test the fix, the Schedule window did not open.
Error message
Error code 4 WLnewPatientId Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: cd /usr/g/hisris more wlsErrors Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.