Have reset gantry, but no Ventri gantry motion.
Symptom
Have reset gantry, but no Ventri gantry motion. Also noted the position on monitor shows 0 degrees but gantry is rotated to about 90 degrees
Error message
ERROR_GCU_POSITION_FAULT (Error #1394) axisID = 7 ROTATION ERROR_GCU_POSITION_FAULT:.:Last axis movement failed.:.:Perform axis extended initialization and retry the operation.:.: Movement position fault description: axisID = 7 severity = 4 motionState = 0 time = 238415130 sourceMsgID = 678058 errorDescription = ax[7] failed SN=8105551.:.:Trace:{1}
ERROR_GCU_HOME_WRONG_POSITION (Error #1329) axisID = 7 ROTATION ERROR_GCU_HOME_WRONG_POSITION:.:Internal motion software error:.:Enter Gantry Handler utility and initialize the axis specified above.:.: Position Error - axisID = 7 severity = 4 motionState = 0 time = 238414851 sourceMsgID = -1 errorDescription = ROTATION-home found in the wrong place expected -56500 found -77723 delta allowed 5000 :.:Trace:{1}Applies to
- Product
- Ventri Ventri Ventri
- Component
- 5321445 (CBL500040) VENTRI Rotation optical limit switches and cable 5321445 (CBL500040) VENTRI Rotation optical limit switches and cable 5321445 (CBL500040) VENTRI Rotation optical limit switches and cable *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Muga patch FMI 40788 when loaded does not indicate it completed
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Part number needed for Optima E-Stop
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- Need part numbers for the Millennium MG/MC table E-Stop