GE Healthcare · MRSigna2225748

RF calibration not in specification.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225748 Analogic RF Amplifier 1 Peak Power Fault Error: 2225748 Host: lx-mr Process: NSP File: HPC PORT ID:spi rf mon Line: 3066 Analogic RF Amplifier 1 Peak Power Fault. Peak power from Analogic RF Amplifier 1 exceeds specification(69.5-69.8dbm). Possible causes: RF Interface (RFI) calibration, defective RFI output stage, defective Exciter or defective RF Amplifier 1

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR