Symptom
Scanner won't boot to applications. Scanner won't boot to applications
Error message
Error: 230011079 Error detected closing a scan file header on the scan data disks. See LD for complete error 230011079 Error detected closing a scan file header on the scan data disks Mon Nov 8 11:15:12 2004 230011079 6 hmaich_gels_ct4 scanmgr sdio_xfs_io_ops.c 1094 Function: Data Acquisition : SBC Processing Scan Type: Helical Scan: 6826/2/1 Scan Seq Id: 4000 Error detected closing a scan file header on the scan data disks. Scan: CT20/6826/2/1 Descriptor: 33 Unix system error : 98 --- Protocol wrong type for socket. Some views may be unavailable for reconstruction until the system is restarted. Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5114536-11 (replaces 5114536-10) Scan Disk Array, Hard drive with mounting sled, Seagate ST37345LC FW001 See Long Description Part #: 5114536-11 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard Drive with mounting sled, Seagate ST373455LC Firmware Revision 001 FRU: Yes Status: Orderable Comments: Per ECR 2049162, 212008898, 10/25/07. This part replaces 5114536-10. See the attachment for further details. Part #: 5114536-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard drive with mounting sled, Seagate ST373454LC Firmware 002 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041216 __Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 See this link to service note for details this part being obsolete and changed up to 5114536-11 : http://www.ct.med.ge.com/tech_pubs/2360027/sn/SN07102512a.pdf
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.