Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230019002230011068

LightSpeed 5.X (80/100kW)

Symptom

Scanner fails to boot to applications, Unable to verify DARC system startup. The system will not boot up to applications with a pink attention box: Unable to verify system functionality on startup. Please power cycle the console. If the system fails to startup after the power cycle, then please call GE Service. FE or OLE sees that they can telnet to the DARC but they cannot ping or rsh with the DARC. If the DARC is observed while booting, Drive X adapter X predicting future failure Press any key to continue is seen.

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Error: 230011068 Unable to initialize scan data disk handling See long description for full error Thu Jun 5 07:32:20 2008 Error: 230011068 Host: wct2 Process: scanfilemgr File: sdio xfs api.c Line: 3561 Function: System Control : Scan Database Manager Scan Type: Not scanning Scan: 0/0/0 Unable to initialize scan data disk handling. Unable to close scan data root directory on scan data disks during start-up processing. Directory: /raw data OS error : 0 --- Success

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5144058-500 Upgrade kit - GOC3 to GOC4. DARC2 is not included. Order proper DARC2 separately - REFER to SN11063012 See Long Description for Details. 5144058-500 Upgrade Kit GOC3 to GOC4 - without DARC2 Part #: 5144058-500 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Service Kit - SDDA and GOC3 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL FRU: Yes Status: Orderable Comments: ** FE ASSIST ** This part number no longer contains a DARC2 it must be ordered in addition per Your system type. REFER to SN11063012a(July 26 2011 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.