Pular para o conteúdo principal
GE Healthcare · CTLightSpeedBrightSpeed185111

LightSpeed 4.X - Octane

Symptom

Unable to tilt the gantry. Unable to tilt the Gantry

Error message

Error: 185111, HW error: Tilt Interference/Limit switches are INVALID See LD for complete error Mon Nov 28 14:39:35 2005 Error: 185111 Host: ETC Process: Control File: etcValidIntfMatrix.c 1.8 7/9/ Line: 133 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 1066 Log Series: 1 HW error: Tilt Interference/Limit switches are INVALID. Interference/Limit switches = 0xd2db. BCK LIMIT: 1 FWD LIMIT: 0 FWD INTF: 0 BCK INTF: 0

Applies to

Models
LightSpeed, BrightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.