Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001708650055153948154175007852005000

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Circular artifact on scan images. Artifact may look like a donut or have light and dark bands. See Long Description for details Circular artifact on scan images. Artifact may look like a donut

Error message

Error: 230017086 DAILY IQ CHECK THRESHOLD FAILURE See Long Description for the full error Error: 230017086 Host: mcop_sbc Process: accs File: aircal_processing.c Line: 797 DAILY IQ CHECK THRESHOLD FAILURE - Scan 50055/8/3 Drift check fail Kv - KV_80, Sfov - SFOV_LARGE, Colimation - SLICE_COLLIMATION_4x125, Focal spot -FOCAL_SPOT_SMALL, Das Gain - Value:30, Row - 0, Channel - 467, Baseline mean - 0.153948, Current mean - 0.154175, Smoothed ratio - -0.007852, Lower limit - -0.005000, Upper limit - 0.005000 .

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2336885 MDAS 4 or MDAS 8 Convertor Board, replaces 2322994-3. See Long Description for details. 2336885 MDAS 4 or MDAS 8 Convertor Card, replaces 2322994-3 (see below) Part #: 2336885 Part Type: GP CONSUMABLES Part Weight: 0.48 LBR Repairable: No Returnable: No Hazardous Material: N Description: SERVICE, MDAS CONVERTOR BD FRU: Yes Status: Orderable 4 Slice versus 8 Slice Configuration The HiSpeed QX/i system can operate 4 slice per rotation, data collections mode, whereas, the Detector and MDAS are capable of operating 4 slice or 8 slice per rotation. From a hardware perspective the main difference is in the MDAS. The 4 slice configuration uses a depopulated MDAS. This means that all EVEN slots are empty. Only ODD numbered slots contain converter boards. Data collection is still performed in the same manner using only the odd converter card slot assignments.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.