GE Healthcare · CTLightSpeedVCT200280012
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Mouse changes state from arrow icon to hour glass icon in viewport. See Long Description for details Mouse changes state from arrow icon to hour glass icon in viewport. Intermittently the mouse changed states while in the viewports, causing havoc in trying to execute mouse clicks.
Error message
Error: 200280012, Attempt to reenter real time mouse routine See Long Description for details. Error: 200280012, Attempt to reenter real time mouse routine Wed May 30 00:58:05 2007 Error: 200280012 Host: stmarysctls Process: afViewer File: ctBoard.cxx Line: 192 Attempt to reenter real time mouse routine
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. 2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. Should NOT be used on GOC2 and above or GRE consoles which contain the HP computer. The HP computer uses a 3 button mouse which is part # 2368697-2 or part # 2348405-2. 2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. Should NOT be used on GOC2 and above or GRE consoles which contain the HP computer. The HP computer uses a 3 button mouse which is part # 2368697-2 or part # 2348405-2. Part #: 2169940-28 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCTANE MOUSE FRU: Yes Status: Orderable Comments: FLS - OCTANE MOUSE & INDIGO COMPUTERU/O NEXGEN 12/97 GWENN MS08992REPLACES 2114561-12 FOR (LX) LIGHTNING. PER PCN, 197770 AND FMI#60476 MUST ORDER 2187507 FOR CONTOUR SAME AS 2142759 (ALL GONE!), WHICH WAS PHASED OUT WHEN STK.WAS USED UP. MS7093 1-7-00 MOUSE AND PAD PER CHIP NLA AS HOTLINE/OBTAINFROM V/C 627584 5/1/00 MOUSE PAD IS 2252872. FE/CUSTOMER NOTE: CT ENGINEERING HAS AUTHORIZED S.G.I. TO DELIVER EITHER LIGHTOR DARK GRAY MOUSE TO FILL ORDERS FOR THIS ITEM. MS7093 HQT: M3 NEXGEN FW 10 '98 * NO LONTER STK'D @ MR MFG'RG 4-00 FOR EXCITE II SYSTEM, USE P/N 2368697-2, PER RSPL OF SAME NAME. MS20544 8/25/03. GELS - SAME AS 2142759,OCTANE MOUSE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.