Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed18310150036

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183101 Two consecutive mA samples out 10.0 % or 4 mA tolerance. See Long Description for full error Error: 183101 Two consecutive mA samples out 10.0 % or 4 mA tolerance. Tue Mar 21 07:18:23 2006 Error: 183101 Host: OBC Process: Generator File: gen tech.c 1.34 Line: 574 Function: XRAY GENERATION : MA Loop Scan: 50036/10/2 Type: Axial Exception Level: Pri/Most Severe Time: 07:18:23:000 Log Series: 235 Two consecutive mA samples out 10.0 % or 4 mA tolerance. Selected technic: 80 kV, 300 mA Sample1 = 0.7 mA. Sample2 = 0.1 mA.

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2208852 - HP INVERTER RHAPSODE - used on CT/i, 1.x, 2.x, 3.x and 4.x systems. 2208852 HP INVERTER RHAPSODE 2208852 HP INVERTER RHAPSODE Part #: 2208852 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP INVERTER RHAPSODE. FRU: Yes Status: Orderable Comments: IF RHAPSODE SYSTEM CURRENTLY HAS T3202JK CAN USE 2208852 . THE LONGER BOLTS NEEDED FOR THE 2208852 INVERTER ARE NORMALLY SUPPLIED IN A BAG WITH THE INVERTER OR CAN BE ORDERED SEPARATELY UNDER P/N 46-208561P43
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues