Pular para o conteúdo principal
GE Healthcare · MRSigna22363802236568

CERD DIAGS and CERD BOOT Errors Logged on Every TPS Reset and System Boot Session due to UNIVERSAL CERD DIGITAL Board.

Symptom

CERD DIAGS and CERD BOOT Errors Logged on Every TPS Reset and System Boot See Long Description The FE and customer have noticed that on every system boot up and on every TPS reset that the two errors shown below are consistently logged. There appears to be no effect on the system, the TPS resets are successful, but these errors shouldn't be happening and the customer wants it fixed. The FE, of course, agrees. Tue Jul 1 07:33:10 2008 2236380 2 ALAMOMRI NSP TPS:aimed_ccrcci.c 382 CERD DIAGS - AIME CCR/CCI Test Failure. The CCR/CCI test failed using alternating CCI assertions. The failure is in the imaginary component of the results. Address: 00000003H Data exp: 000008H Data read: fffffffeH FRU(s): CERD Digital Board Tue Jul 1 07:33:10 2008 2236568 2 ALAMOMRI NSP TPS:prgasc.c 296 CERD BOOT ERROR MESSAGE - ON CAP - ASC DIAGS ERROR Failed while running asc power up diags. Status: ffffffffH

Error message

2236380 CERD DIAGS AIME CCR/CCI Test Failure. The failure is in the imaginary component of the results. See Long Description 2236380 2 ALAMOMRI NSP TPS:aimed_ccrcci.c 382 CERD DIAGS - AIME CCR/CCI Test Failure. The CCR/CCI test failed using alternating CCI assertions. The failure is in the imaginary component of the results. Address: 00000003H Data exp: 000008H Data read: fffffffeH FRU(s): CERD Digital Board

2236568 CERD BOOT ERROR MESSAGE - ON CAP - ASC DIAGS ERROR Fri Feb 25 14:26:22 2000 Error: 2236568 Host: lx-mr Process: NSP File: TPS:prgasc.c Line: 281 CERD BOOT ERROR MESSAGE - ON CAP - ASC DIAGS ERROR Failed while running asc power up diags. Status: ffffffffH

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Universal CERD Digital 2206099 See Long Description This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues