CWA ID109238. Will Not Scan with Gradient Errors.
Symptom
Will Not Scan with Gradient Errors.
Error message
2218527 Scanning has stopped. Gradient amplifiers are not ready. . Mon Jun 16 11:29:36 2008 Error: 2218527 Host: lx-mr Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. Gradient amplifiers are not ready 2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.
Applies to
- Models
- Signa, Discovery, Optima
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution