Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118267 (90-0706) Wrong DSP version from - CT TIC See LD for error explanation Code 90 Applications Errors Error code: 90-0706H Message/explanation: Wrong DSP version (Engineering debug) Potential cause: PPC kV control board Recommended action/ Troubleshooting guide: Reload the software and database If no improvement, change PPC kV control

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5136161-4 PPC KV Control replaces (5136161, 5136161-2, 5117564-10) used by all 5.X, Brightspeed and VCT gantries See Long Description for Details PPC KV Control Bd. 5136161-4 replaces (5136161, 5136161-2, 5117564-10) this is the orderable Programmed KV Control Board for HP60, LSPro 80KW , 100KW and VCT 7.X, RT16, BrightSpeed Power and Select Systems (Velocity, Delight, Meri) Part #: 5136161-4 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL P6.33BF6 AS FRU FRU: Yes Status: Orderable Current: 5136161-4 Harvested: Not Available Exchange: Not Available Comments: Jedi KV Control Board used on Definium 8000, Revolution XR/d-2X (Lightning), Definium 6000, and Precision 500D. Also may be used on all Innova systems Bolus/Unity/SPAD/Harmony. P.Liebau 1/2010 Part #: 5136161-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL 6.30BFA AS FRU FRU: Yes Status: Orderable Comments: In order to avoid Jedi software download issues on Definium 8000, Revolution XR/d-2X (Lighting), Definium 6000, and Precision 500D you may order 5136161-4, KV Control Board developed for these systems. Part #: 5136161 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED KV CTRL P6.30BFA as FRU FRU: Yes Status: Orderable Comments: (2/21/06 212001261) The 5136161 KV board replacement is the preferred part used on all HP60 hp 80 LS-RT and VCT products.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.