Pular para o conteúdo principal
GE Healthcare · CT1209963204800096000009690796007

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 120996 The High Speed Disk failed to read/write known data. Scanning stopped. See long description for full sample error. Mon Apr 21 08:21:20 2008 Suite: SVMC Host: SBC0 Proc: scse Error: 120996 File: ScSeSet.m 1.47 Method: isScseReady Line: 2952 Function: SCANNING : Hardware Preparation The High Speed Disk failed to read/write known data. Scanning stopped.

Error: 0 shutting down local host | due to startup failure: sdCoord See long description for full sample error. Mon Apr 21 08:34:25 2008 Suite: SVMC Host: SBC0 Proc: hostMonitor Error: 0 File: log_it.c 1.12 Method: log_it Line: 89 shutting down local host | due to startup failure: sdCoord

Error: 32048 AP ERROR(072:811:00096:00000) IOPB Error. See error code. drive faulted See long description for full sample errors. Mon Apr 21 08:21:20 2008 Suite: SVMC Host: SBC0 Proc: scse Error: 32048 File: MaxiDisk.m 1.8 Method: devRead:length: Line: 422 AP ERROR(072:811:00096:00000) IOPB Error. See error code. drive faulted Mon Apr 21 08:34:23 2008 Suite: SVMC Host: SBC0 Proc: sdCoord Error: 32048 File: MaxiDisk.m 1.8 Method: devRead:length: Line: 434 AP ERROR(072:569:00000:00000) No message defined for this AP error.

Error: 0 Proc

Applies to

Product
HSA Z Series HSA Z Series HSA Z Series Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-313030G1 (T3202CR) High Speed Sabre Disk Drive used on HSA Z series systems See Long Description for details Part #: 46-313030G1 or T3202CR Part Type: GP REPAIRABLES Part Weight: 19.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: SABRE ASSY.// T3202CR FRU: Yes Status: Orderable Comments: FLS - PER M. GRABOWSKI LOW SHARE PRICING II 12/94 KJM
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.