Pular para o conteúdo principal
GE Healthcare · MRSigna223919022391832239191

CWA ID108974. Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2239190 GRADIENT DRIVER TESTS - Power Up MIF Z Axis Register Addressing Test Failed. See long description Error: 2239190 Host: GELXMR93 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 3337 GRADIENT DRIVER TESTS - Power Up MIF Z Axis Register Addressing Test Failed. Possible causes: cables within the GAP module, cable between MIF & GAP, the MIF board, the GAP board, or the MIF DIP switches are improperly set.

2239183 GRADIENT DRIVER TESTS - Power Up Register Pattern Test Failed. See Long Description. Error: 2239183 Host: GELXMR93 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 3335 GRADIENT DRIVER TESTS - Power Up Register Pattern Test Failed. Each ASM board has several errors. Possible causes: GAP board or GAP +12V Supply.

2239191 GRADIENT DRIVER TESTS - Power Up ASM Y Axis Register Addressing Test Failed. Fri Mar 27 14:49:51 2004 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2239191 File: PMH.m 1.29 Method: gap Line: 3337 GRADIENT DRIVER TESTS - Power Up ASM Y Axis Register Addressing Test Failed. Possible causes: cables within the GAP module, cable between MIF & GAP, the MIF board, the ASM board, or the GAP board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution