Can't scan with multicoils due to failing SSM.

Symptom

Unable to scan with multicoils with error 2225694 & 2225692. Head and body mode works fine. Unable to scan with multicoils with error 2225694 & 2225692. Head and body mode works fine.

Error message

2225692 MDS link failure. A receive error was detected on an MDS response. Check the connections and signal level on the MDS fiber optic link. See Long Description Sun Jun 22 00:43:27 2008 Error: 2225692 Host: LX-MR Process: NSP File: HPC_PORT_ID:spi_mds Line: 502 MDS link failure. A receive error was detected on an MDS response. Check the connections and signal level on the MDS fiber optic link. Timer: 7 UART status: 0x5B Received byte: 0x05

2225694 MDS link failure. A response packet was lost due to a UART receive error. See Long Description Sun Jun 22 00:42:51 2008 Error: 2225694 Host: LX-MR Process: NSP File: HPC_PORT_ID:spi_mds Line: 504 MDS link failure. A response packet was lost due to a UART receive error. If the link is broken, a separate error message will be logged. Timer: 33 First response byte: 0x0A Second response byte: 0x60

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2160200-43 SSM for LX See attached file for part information and Engineering Documentation.
Area
MR