Exciter Calibration Errors And Power Monitor Trips only when using T/R knee coil. due to defective CERD exciter.
Symptom
Exciter Calibration Errors And Power Monitor Trips Intermittent exciter calibration errors and 85 faults, FE performed power monitor functional checks and they passed.
Error message
2225739 RF Interface (RFI) High Voltage Relay Fault. Thu Mar 7 16:22:42 2002 2225739 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3066 RF Interface (RFI) High Voltage Relay Fault. Possible causes: RF Monitor Cable to System Support Module (SSM) is disconnected or defective. Note: This fault may occur in conjunction with a Power Monitor Fault or when power cycling the SSM.> 2234496 No or little signal amplitude detected during Exciter calibration. Fri Jan 30 16:21:12 1998 Error: 2234496 Host: oc001 Process: NSP File: TPS:cie_exc_cal Line: 455 No or little signal amplitude detected during Exciter calibration. Possible hardware problem - please run CERD Diagnostics! 2225737 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Thu Mar 7 16:22:41 2002 2225737 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 40 pulse width: 33 duty cycle: 03 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when pow
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2148300-8 1.5T CERD Exciter 2148300-8 1.5T CERD Exciter 2148300-8 1.5T CERD Exciter
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.