CWA ID10859. MRCustomerCall Image Quality OVERRANGE IMAGE ARTIFACT.
Symptom
MRCustomerCall Image Artifact Library Image Quality This issue has been written to assist in supporting remote Customer calls for Image Quality issues. See long description for more details. This issue has been written to assist in supporting remote Customer calls for Image Quality issues. Take a look at the attached image gallery and see if the issue matches one of the common problems. Once you identify the name of the IQ issue, you can go to the resolution concept titled by that name to get information on that type of image quality problem. Additionally, it is good to ask the following general questions about the issue to help understand. Artifact Characterization/History of Occurrence Questions •How long has artifact been occurring? •Frequency of occurrence, often or intermittent? •Check Chronology history any images previously pulled? •Patient Specific or Multiple Patients? •Coil Specific or Multiple Coils? Have they reseated the coil and re scanned?
Error message
2234211 MR Signal too large. Reduce Receive Gain 1. See Long Description Error: 2234211 Host: MR1SMR Process: NSP File: TPS:asc_check_RAP_overrange Line: 206 MR Signal too large; Reduce Receive Gain 1.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.