Intermittent Scan Aborts when table is moving between 1st and 2nd scouts, between helical series, or between axial series.

Symptom

Intermittent Scan Aborts when table is moving between 1st and 2nd scouts, between helical series, or between axial series. Clicking Pause / Resume restarts scan. See long description for details. Intermittent Scan Aborts when table is moving between 1st and 2nd scouts, between helical series, or between axial series. The customer can click "Pause" and then "Resume" and continue scanning, after clicking "Resume or "Pause" it can take from 10 to 60 seconds before system will start scanning again.

Error message

Error: 184189 SW ERROR: Software Problem. Error occured when writing CF message 10021. See LD for complete error Mon Jul 11 09:51:24 2005 Error: 184189 Host: STC Process: Scan Control File: scRxMgr.c 1.48 Line: 251 Function: No System Function Reported < Scan: 12469/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:51:23:970 Log Series: 958 CF error 2103 detected on sending a CF message. SW ERROR: Software Problem. Error occured when writing CF message 10021

Applies to

Models
HiSpeed, VCT
Product
HiSpeed QX/i - GRE Always MDAS 4 QX_P_M4 HiSpeed QX/i - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an MDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. xxx
Component
2245480 SIGNAL BRUSH TIP SET FOR HSDCD See Long Description for details Part #: 2245480 Part Type: GP CONSUMABLES Part Weight: 10.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSH TIP SET FOR HSDCD FRU: Yes Status: Orderable Comments: FLS - SIGNAL BRUSH TIP SET FOR HSDCDHINO LXI,DXI,FXI,ZXI/CTE MS08992 3/7/00 FE NOTE: THIS KIT CONTAINS 40 BRUSHES WITH CAPS. MS03185 11/02/01 DO NOT DISCOUNT - PER SHAKUN 32 PCS 2253677 MAT&apos;L CA40 TIPS 8 PCS 2253678 MAT&apos;L CAG TIPS 40 PCS 2253686 BRUSH CAPS ON PROMOTON THROUGH JUNE 30, 2004-DO NOT CHANGE PRICE PER SHAWN (MAYND) GELS -
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.