Millennium gantry motion initialization fails after AUI boot Millennium gantry motion initialization fails to…
Millennium gantry motion initialization fails after AUI boot
Symptom
Millennium gantry motion initialization fails after AUI boot Millennium gantry motion initialization fails to start after AUI boot. No message appear for asking for motion initialization No communication between Millennium Acquisition GIP software and 4-Axis motion controller Millennium gantry motion initialization fails to start after AUI boot. After application boots no message appear for asking for motion initialization. Under service login, FSE try to initialize MC without success. On log file we can see gip_error: GIP error: Time out on controller response Last message sent to controller by gip was STPG<00><00><00><10>" No communication between Millennium Acquisition GIP software and 4-Axis motion controller Contributor 100008553 Certo July 2, 2008
Error message
GIP error: Time out on controller response Last message sent to controller by gip was STPG " GIP error: Time out on controller response Last message sent to controller by gip was STPG<00><00><00><10>" No communication between GIP and 4-Axis MC.
Applies to
- Models
- Millennium
- Product
- MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT MPR/MPS Millennium MPR Millennium MPS
- Component
- *IPS IPS IPS
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Need part numbers for the Millennium MG/MC table E-Stop
- How do I Delete a file on Genieacq that the system thinks is still acquiring
- VG, the contour works... but in the opposite way.
- MG_MC_MyoSight_MPR_MPS Bar Phantom Resolution Size