GE Healthcare · MRSigna22185272217995
CWA ID107525. Unable to scan with message Prescription allocation request fail.
Symptom
Unable to scan with message Prescription allocation request fail System is able to prescan. But when the prescan ends and the scan is about to begin, it fails with a message "Prescription allocation request failed". This message doesn't appear in the log, but another one does (see below): Error: 2218079 Host: MR6 Process: msgView File: ImageStat.cc Line: 68 Image space query failed
Error message
2218527 Scanning has stopped. not enough memory available to start acquisition, please try again. 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again. Error: 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again. 2217995 Image file allocation failed. Please recover space for X images. See Long Prescription Error: 2217995 Host: UPCH-MR Process: ifcc File: IfccScan.cc Line: 3529 Image file allocation failed. Please recover space for 21 images.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.