Pular para o conteúdo principal
GE Healthcare · MRSigna22415352241534

Gradient Data Overrange Occurred.

Symptom

Will Not Scan with Gradient Errors.

Error message

2241535 Y Gradient Data Overrange Occurred. The GIP detected an error in the incoming High Speed Data from the IPG. Tue May 6 15:52:51 2008 Error: 2241535 Host: GE_MRI Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2586 Y Axis Gradient Data Overrange Occurred. The GIP detected an error in the incoming High Speed Data from the IPG. Swap fiber optics at the Systems Cabinet. Run fiber optic Diagnostic

2241534 X Axis Gradient Data Overrange Occurred. The GIP detected an error in the incoming High Speed Data from the IPG. Tue May 6 14:44:12 2008 Error: 2241534 Host: GE_MRI Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2586 X Axis Gradient Data Overrange Occurred. The GIP detected an error in the incoming High Speed Data from the IPG. Swap fiber optics at the Systems Cabinet.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2160818-7 IPGII Board SUN STYLE FRONT PANEL See Long Description IPG2 Board 2160818-7 FEMC Part Description: IPGII SUN STYLE FRONT PANEL
Area
MR

Full cause & step-by-step resolution