Pular para o conteúdo principal
GE Healthcare · X-ray R&FDefinium

Def5000 The gantry elevation will only drive in low speed.

Definium 5000 The Def 5000 will only drive the gantry elevation in low speed due to an improperly miss adjusted Height Inverter.

Symptom

Def5000 The gantry elevation will only drive in low speed. The Def 5000 will only drive gantry elevation in low speed. Any high-speed drive or auto-positioning attempt of elevation will result in immediate halt of gantry and an error displayed on the touch screen “FATAL ERROR – HEIGHT POTENTIOMETER ERROR”. A system reset is required to recover. No error is reported in error log of Touchscreen or system error log viewed from SUIF on magic PC.

Error message

Def5000 FATAL ERROR – HEIGHT POTENTIOMETER ERROR Error was not actually in the error log, but the Touch Screen displays "FATAL ERROR – HEIGHT POTENTIOMETER ERROR".

Applies to

Models
Definium
Product
Definium 5000
Component
Def5000 Positioner Height Inverter
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required