Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000237750619

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Customer can not scan with DAS temperature errors. Customer can not scan with DAS temperature errors

Error message

Error : 260002377 Das Temperature is HIGH See long description for full error Mon Jun 2 10:20:59 2008 Error: 260002377 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 752 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50619/3/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:20:59:205 Log Series: 86 DAS Temperature is HIGH. Highest temperature was seen is on Converter Board 25. Measured: 62.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. Converter Board Bit Map: 0x000001000000. Bit Map Indicates Failed Board(s) for Multiple Board Errors. (Board Bit Map defines: BIT47 = CnvBrd 48 ... BIT0 = CnvBrd 1)

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5363182 (replaces 46-220248P3) H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU and SDAS See long description for details Part #: 5363182 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR 2089514, 212008898, 02 Nov 2009. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5363182 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2089514 46-220248P3 FAN: TUBE AXIAL, 115CFM, 115V, 60HZ, Used on H1 and H2 gantries for SDAS Fan / Plenum assembly Part #: 46-220248P3 Part Type: GP CONSUMABLES Part Weight: 1.50 LBR Repairable: No Returnable: No Hazardous Material: N Description: FAN: TUBE AXIAL, 115CFM, 115V, 60HZ FRU: Yes Status: Orderable Comments: FLS - BALL BEARING FAN NECESSARY FOR ROTATION ON CT GANTRY USED ON HELIOS 1 AND HELIOS 2 PER PCN 199761 MS08992
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues