GE Healthcare · X-ray R&FDefinium
Def5000 The generator restore from backup CD never completes.
Definium 5000 The generator restore from backup CD never completes due to an incorrect switch setting on the ATP Console Board.
Symptom
Def5000 The generator restore from backup CD never completes. When attempting to restore the Def 5000 generator from the backup CD, the positioner touch screen displays a backup/restore screen with the progress bar indicator stuck at approximately 10 % complete, and the Magic PC displays GENERATOR RESTORE IN PROGRESS. The restore never completes.
Applies to
- Models
- Definium
- Product
- Definium 5000
- Component
- Sedecal Generator
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.