Pular para o conteúdo principal
GE Healthcare · CT00000532

ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt)

Symptom

Unable to scan with XG errors

Error message

XG ERROR 117 with a 00000532 See Long Description for details ER–117 00000XYY XG Error Error Code X = 0: Test Mode 1: Stand–by 2: Preparation Acting 3: Generator Ready 5: Exposure 7: Transition Step to Stand–by Error Code YY = 01:No 550 VDC 02:15V_Safe Error 03:Over Volt Capacitor 11: Rotor Controller Error 12:Rotor Acceleration Error 013: Rotor Feedback Test 014: Rotor Fault Detect Test 16:Low Current Error 18:Unbalance Error 21:Filament Overcurrent 021: Feedback Test (X=0) 22:No Filament Current 022: Fault Detection Test (X=0) 23:Bad Relay Return 023: Small Focus Test (X=0) 24:No Filament Voltage 024: Large Focus Test (X=0) 25:Filament Control Error 526: MA Measurement Error (X=5) 031: Feedback Test (X=0) 032: Fault Detection Test (X=0) 033: Close Loop Test 531: MA Accuracy Error 532: KV Accuracy Error (X=5) 533: Spit Error (X=5) 534: Exposure Backup Timer Error (X=5) 535: Anode Over mA (X=5) 536: Cathode Over mA (X=5) 537: Anode Over kV (X=5) 538: Cathode Over kV (X=5) 539: Anode Shoot–through (X=5)

Applies to

Product
ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2205389 Non EMC Anode Tank - used on early CTi and Prospeed 2205389 Non  EMC Anode Tank PLEASE use T3109CD/2203602 FOR SUB-SECOND SCANNERS and IF the SYSTEM USES the PERFORMIX TUBE 2205389 Non – EMC Anode Tank PLEASE use T3109CD/2203602 FOR SUB-SECOND SCANNERS and IF the SYSTEM USES the PERFORMIX TUBE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues