GE Healthcare · CTLightSpeedVCT230023053244633
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cradle unlatches on its own and the scanning aborts
Error message
Error: 230023053 Failed on Gantry Firmware Mon Feb 4 07:19:52 2008 Error: 230023053 Host: aafgect1 Process: dailyPrep Rx File: DailyPrepUtils.c Line: 5433 Failed on Gantry Firmware. Cannot continue. Error: 244633 Table reports that Table Communication becomes down. Wed May 25 17:35:00 2005 Error: 244633 Host: ct01 Process: fwmgr File: fwmgr_events.c Line: 7301 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Table reports that Table Communication becomes down.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2202215-2 Emergency switch See long description for details Part #: 2202215-2 Part Type : GP CONSUMABLES Part Weight : 0.09 LBR Repairable: No Returnable: No Hazardous Material : N Description: SWITCH FRU : Yes Status: Orderable Comments: FLS - EMERGENCY SWITCH ASM U/O CT HISPEED-YMS GANTRY M3 FW20 '98 MS1101 3/98 USED ON SMART GANTRY. SOURCED BY ASIA. MS20544 ON 7/10/02. Linkage of 2202215-2 -> 2388230 can not be setup, pls order 2388230 if zero on hand in GPRS for 2202215-2. CN3218 on 7/19/06
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required