P500D Aurora install startup issue
Symptom
P500 Aurora install startup issue: Error 1090115 (OTS I/O Node Guarding) in log. Operational problems: lat and long OTS locks won't release, vertical and angulation OTS locks free floating, OTS self test fails, Collimator alive test pass, Can Bus resistance and voltage checks pass. On OTS I/O board, test points for voltages failed for TP9=.6V(5.5V), TP17=.3V(5V), TP11= 1.5V(6V), TP12=0V (3.3V). FE had already replace I/O board assy.
Error message
P500D OTS IO node guarding error (1090115). 1090115 Recovery Class: WARNING Error Text: OTS IO node guarding error. Major Function Impacted: Pos Minor Function Impacted: DeviceGroup View Level: opr Root Cause: 1 Description: OTS IO is not responding. Indicates a CAN device or bus problem. Refer to the CAN bus block diagram. - If a single device is reporting an error, check that device, its power, and its cabling/connections. - If multiple device are reporting errors, check common cabling/connections and common power. - If all CAN devices are reporting errors, check common cabling/connections. Perform OTS, and CAN Bus and IO Board Diagnostics, reset system. If problem persists, replace IO Board. Also check/ replace the System Controller. Recommended Action(s): Retry or reset the system Customer Text (English): OTS I/O not alive. Retry or reset the system.
Applies to
- Software
- Aurora
- Component
- OTS
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- AEC Backup Time message
- New style ion chamber does not have a preamp
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug