Pular para o conteúdo principal
GE Healthcare · MRSigna2238454

2238454 Power Module Poly Phase Buck Module Error due to failed power module.

Symptom

Will Not Scan with Gradient Errors.

Error message

2238454 % axis LOW 8645 Power Module Poly Phase Buck Module Error occurred. Sat Mar 22 12:56:57 2008 Error: 2238454 Host: GVRHMR01 Process: NSP File: HPC PORT ID:GAP/GIP/GP Line: 1038 Z axis LOW 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM. Sat Mar 22 12:56:57 2008 Error: 2238454 Host: GVRHMR01 Process: NSP File: HPC PORT ID:GAP/GIP/GP Line: 1038 Z axis LOW 8645 Power Module Poly Phase Buck Module Error occurred. Possible causes: high line volt. or overtemp. Wait a few minutes for cooling, then turn 8645 power off for 30 seconds to clear error. After 30 seconds, turn 8645 power back on. Try to scan. If Main Control Board LED is lit again, replace PPBM.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-328206P115 CROWN A10999-G2 8645 PWR MOD W/CROWB
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required