Pular para o conteúdo principal
GE Healthcare · MRSigna2225385

Replacement SCIM Causes WIM Initialization Error due to Wrong Type SCIM Installed.

Symptom

Replacement SCIM Causes WIM Initialization Error See Long Description Upon replacement, new SCIM causes WIM Initialization error (See below). APS and MPS fails prescan. 2225385 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_cpb_initialize_fail Line: 4 Workspace Interface Module Initialize failed. If recovery fails, check Workspace Interface Module connections and run Workspace Interface Module diagnostics.

Error message

2225385 Workspace Interface Module Initialize failed. See Long Description Mon May 23 08:17:18 2005 Error: 2225385 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_cpb_initialize_fail Line: 4 Workspace Interface Module Initialize failed. If recovery fails, check Workspace Interface Module connections and run Workspace Interface Module diagnostics.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2289697-2 SCIM
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.