GE Healthcare · MRSigna2225553
Cannot Scan With intermittent TR and Dynamic Disable Faults.
Defective coil.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2225553 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 3. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225553 File: PMH.m 1.29 Method: spi_tr_mon Line: 6250 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 3. Check cable to multi-coil board or check for an open pin diode.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2215261 1.5T Torso Phased Array Flex Coil See long description for Coil Shipping list and part numbers. This information is Property of GE Healthcare - For GE Healthcare Service Personnel Only. Disclosure to Third Parties Prohibited. From the SIGNA® TORSO PHASED ARRAY COIL REV 2 DIRECTION 2218805 Documentation. To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS . 1.5T TORSO PHASED ARRAY COIL SHIPPING LIST
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.