VGH system will not setup for scan get message Radial Lateral axis not moving, will not resume.

Symptom

VGH system will not setup for scan get message Radial Lateral axis not moving, will not resume. VGH system will not setup for scan, get message Radial Lateral axis not moving, will not resume.

Applies to

Models
Millennium
Product
VGH VG and VG Hawkey and Varicam
Component
5197979 (PCA000492) MCH Board for SPARE PART 5197979 (PCA000492) MCH Board for SPARE PART 5197979 MCH Board is not compatable with VG motion software version 6.0AH Verify VG Gantry Firmware version. VG Gantry Firmware Version must be 6.0P7 . See KOID 153674 *Gantry Gantry Gantry