Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26000245550723

Scan Hardware Stopped Scan with DCB has an Outstanding SelfTest Failure

Symptom

Scan Hardware Stopped Scan with DCB has an Outstanding SelfTest Failure See Long Description for error string Scan Hardware Stopped Scan message. The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. Data Acquisition System Has Detected An Error In The 5V Analog DAS Power Supply, and DCB Power Supply selftest failed Errors posted in gesyslog whenever a tube warm up was attempted: Tue Dec 7 07:23:22 2004 Error: 245043 Host: ct02_sbc Process: scanmgr File: scanmgr_events.cxx Line: 1595 Function: System Control : Scan Control Scan Type: Cine Scan: 0/0/0 Scan Seq Id: 1000 Scanning System Aborted Scanning Tue Dec 7 07:23:22 2004 Error: 260002455 Host: DAS Process: DCB Firmware File: scanrx.c Line: 290 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/1 Type: Axial Exception Level: Pri/Soft Time: 07:23:22:900 Log Series: 36 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ). Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 4 Tue Dec 7 07:23:22 2004 Error: 260002358 Host: DAS Process: DCB Firmware File: scanrx.c Line: 677 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:23:23:210 Log Series: 37 Scan ABORT -- DAS Has Detected An Error.^B Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) Errors noted previously in the geyslog, prior to the customer attempting a tube warm up. These messages occurred during the prior system boot up: Tue Dec 7 07:20:51 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 07:20:50:115 Log Series: 23 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002403 Notification level: 5 Sequence number: 1 Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 729 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 07:20:50:140 Log Series: 23 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply.^B Measured: 2522 mV, Expected range: 4750 to 5250 mV Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 783 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:240 Log Series: 23 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply.^B Measured: 508 mV, Expected range: -5250 to -4750 mV Tue Dec 7 07:20:51 2004 Error: 260002318 Host: DAS Process: DCB Firmware File: hdwtest.c Line: 110 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:505 Log Series: 23 A Failure Has Been Detected By A DCB Selftest.^B DCB Power Supply selftest failed.

Error message

Error: 260002455 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ) See Long Description for the full error code Tue Dec 7 07:23:22 2004 Error: 260002455 Host: DAS Process: DCB Firmware File: scanrx.c Line: 290 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/1 Type: Axial Exception Level: Pri/Soft Time: 07:23:22:900 Log Series: 36 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ).

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225218 DCB HOUSING ASM for SDAS See LD for details Part #: 2225218 Part Type: GP REPAIRABLES Part Weight: 0.66 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: DCB HOUSING ASM (SDAS) FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2225218 Harvested: Not Available Exchange: Not Available Comments: FLS - DCB ASM, P/N 2225218 7/31/02- C2R COMPLETED PER M DIGGINS, CHANGE ORDER TO T3202KJ BEFORE SHIPPING. MS12464 CAN'T ORDER DAS ASSY SEE T3202KJ 1/2/02 VEN HAS 20 PC MINIMUM BUY PER PO AND 60 DAY LEAD TIME. MS09121 *** CD & S NOTE **** BEFORE REPLACING THE SDAS DAS CONTROL BOARD (DCB), MEASURE THE +5V ON THE BOARD FROM TP-1 - TP-2. VERIFY THIS IS ABOVE4.8 VOLTS. VOLTAGE LOWER THANTHIS MAY CAUSE INTERMITTENT OPERATIONS. IF THE VOLTAGE ISOK ON THE CONVERTER CARD CHASSCHASSIS, YOU MAY HAVE A CABLE PROBLEM. MS30863 5/31/02 GELS - CSO0111459 FRANCE F-45071720002 /04/01/TS FIBER OPTIC FREQ 50 MHZ / NEEDS DIP BOARD 2275604 OR 2245261 PDB 02/01/21,DCB HOUSING ASM
Area
CT