GE Healthcare · CTHiSpeedLightSpeed1831071078518310065001183183960822
HiSpeed QX/i - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183107 KV fault detected: Anode shoot-thru. See long description for complete error Thu Mar 10 12:46:26 2005 Error: 183107 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 239 Function: XRAY GENERATION : KV Loop Scan: 10785/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 12:46:26:065 Log Series: 608 KV fault detected: Anode shoot-thru. Fault latch address = ffcff3 hex, Bit = D2 Error: 183100 Two consecutive kV samples out 10.0 % tolerance Open Long Description to see error example Tue Jun 7 03:35:46 2005 Error: 183100 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 521 Function: XRAY GENERATION : KV Loop Scan: 65001/11/14 Type: Axial Exception Level: Secondary Time: 03:35:45:575 Log Series: 223 Two consecutive kV samples out 10.0 % tolerance. Selected technic: 140 kV, 56 mA Sample1 = 93.919 kV. Sample2 = 0.000 kV. Error: 183183 kV out of 3 % tol See Long Description for error example Error: 183183 Suite: CT01 Host: OBC Proc: Generator File: gen tech.c 1.32 Method: No Method Line: 791 Function: XRAY GENERATION : KV Loop 960822 1230 10922 Scan: 0/0/3 Type: Static Exception Level: Sec/Soft Time: 12:17:46:610 Log Series:
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2208852 - HP INVERTER RHAPSODE - used on CT/i, 1.x, 2.x, 3.x and 4.x systems. 2208852 HP INVERTER RHAPSODE 2208852 HP INVERTER RHAPSODE Part #: 2208852 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP INVERTER RHAPSODE. FRU: Yes Status: Orderable Comments: IF RHAPSODE SYSTEM CURRENTLY HAS T3202JK CAN USE 2208852 . THE LONGER BOLTS NEEDED FOR THE 2208852 INVERTER ARE NORMALLY SUPPLIED IN A BAG WITH THE INVERTER OR CAN BE ORDERED SEPARATELY UNDER P/N 46-208561P43
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.