GE Healthcare · CTHiSpeedLightSpeed182912182904
Gantry table controls would not work when moving patient into the gantry
Symptom
Gantry table controls would not work when moving patient into the gantry. Footpedals would move table up/down.
Error message
Error: 182912 Gantry/Tableside FAST button is non-functional See the LD for the full error code Error: 182912 Gantry/Tableside FAST button is non-functional Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional Error: 182904 A table side key is either stuck or pressed during initialization or just after Emergency Off released. See the Long Description for the full error code Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 345 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2114535 Right Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi See LD for more info 2114535 Right Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi 2114535-2 Left Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi See LD for more info 2114535-2 Left Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi Note: Rubber Overlay for the control panel is 2225299-2
- Area
- CT
Full cause & step-by-step resolution