GE Healthcare · MRSigna2225718
Air Conditioning failure in the Equipment room caused over temperature faults.
Symptom
Unable to scan with RF overtemp faults.
Error message
2225718 TELI RF Amplifier 1 Overheat Fault. TELI Amplifier 1 reports the thermal switch has exceeded 80 degrees C. 2225718 Host: lx-mr Process: NSP File: HPC PORT ID:spi rf mon Line: 3064 TELI RF Amplifier 1 Overheat Fault. TELI Amplifier 1 reports the thermal switch has exceeded 80 degrees C. Possible causes: restricted air flow, blown fan fuse, defective fan, room environment out of specification or defective RF Amplifier 1.-
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR