MG_MC_MyoSight Image shift (jump) between detectors, the image is displaced between detectors in Y direction

Symptom

MG_MC_MyoSight Image shift (jump) between detectors, the image is displaced between detectors in Y direction COR acquisition shows Y mean position: Det 1 (Y) - Det 2 (Y) above acceptable limits </=2.0 Table below shows example COR acquistion test with more than 8mm difference between the Y mean position for det1 & det2. (239.24 - 230.64 = 8.6 mm) COR Acquisition Det 1 Requirement Measurement COR value </=5.0 mm 1.04 mm COR variation </=3.5 mm 0.42 mm Y position 239.24 Y variation </=3.5 mm 1.99 mm Det 2 Requirement Measurement COR value </=5.0 mm 0.95 mm COR variation </=3.5 mm 0.51 mm Y position 230.64 Y variation </=3.5 mm 3.14 mm Mean Y position Det 1 (Y) - Det 2 (Y) </=2.0 8.6

Applies to

Models
Millennium
Product
MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
Component
*Collimator/ Collimator cart Collimator/ Collimator cart
Area
Nuclear Medicine