Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1144566814877214230011036230019002

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

System will not stay up in applications and shutsdown. System will not stay up in applications and shutsdown

Error message

raw_data - lost chains " cross-linked with another chain ( found in the iceConsole.log ) See LD for Error Details To View Scan Data Disk Errors in the iceConsole.log cd /usr/g/service/log tail -100 iceConsole.log | more (tail –500 would show the last 500 lines of this log) (use the space bar to page through this log) Look for any Scan Disk Errors as seen below: "/raw_data - lost chains " cross-linked with another chain "/raw_data - lost chains " cross-linked with another chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a13" has illegal cluster in chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a14" has illegal cluster in chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a18" has illegal cluster in chain

Error: 230011036 Unable to initialize scan data disk handling. see lond description for details Tue Feb 26 22:51:35 2008 Error: 230011036 Host: bmvi Process: queuemgr File: sdio xfs io ops.c Line: 1228 Function: Image Generation : SBC Processing Scan: 0/0/0 Image: 0/0/0 Unable to initialize scan data disk handling. Unable to open information file. Info file: /raw data/info file Mode: 0h Permissions: 0 (octal) OS error : 2 --- No such file or directory

Error: 230019002 

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.