GE Healthcare · Nuclear Medicine889425

When booting Ventri from Standby it skip's the normal system self test sequence, you don't get the message, pull out tab…

When booting Ventri from Standby it skip's the normal system self test sequence, you don't get the message, pull out table clutch and press the psd, it goes right to System Initialization where it cra…

Symptom

When booting Ventri from Standby it skip's the normal system self test sequence, you don't get the message, pull out table clutch and press the psd, it goes right to System Initialization where it crash during rotation. Start up test the "pull table Clutch" do not come up when booting from Stand By. It crashes during Axis Initialization for Rotation, the Rotation starts but halts and messages "Internal Motion Sw Error" appeared. In the problme report you will see: ERROR=Error in GCU .../gcu/application/MotionSynchronizer.cpp(953):: 889425 ROTATION motion is disallowed because LINEAR CAN node is inactive However, if the system is completely shutdown with the breaker and rebooted, it will come up correctly. Contributor 212010832 Oheim 03 April 2008

Error message

ROTATION motion is disallowed because LINEAR CAN node is inactive 14:39:06 ERROR=Error in GCU .../gcu/application/MotionSynchronizer.cpp(953):: 889425 ROTATION motion is disallowed because LINEAR CAN node is inactive

Applies to

Product
Ventri Ventri Ventri
Component
*IPS IPS IPS
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.