Diagnose gantry tilt faults — tilt disabled or noisy during tilt operation

Use this page when the gantry will not tilt, or when it tilts but makes abnormal noise while doing so. The tilt drive on this gantry is a power cylinder + motor assembly at the gantry base — most of the faults below trace back to that assembly, its microswitch, its connector/cable, or the mechanical parts around it (pillow block, gas spring, cover).

Applies to
Activion16 (TSX-031A). Tilt drive = power-cylinder + motor assembly at the gantry base. Mechanical/electromechanical fault chart — no KGTSM error code accompanies these faults in the source.
Steps
0 steps
Source
2D201-112EN*D §1.2.1 Troubleshooting and repair of the tilt section (pp.35–36)

Tools & parts

ItemPart numberQty
Power cylinder (tilt drive assembly — includes the tilt motor and solenoid brake)n/a in source
Tilt-position microswitchn/a in source
Tilt drive unit connectorn/a in source
Transportation (shipping) jign/a in source
Tilt drive interconnect cablen/a in source
Gas spring (tilt frame/cover assist)n/a in source
Pillow block (tilt bearing) unitn/a in source
Front cover (interference point)n/a in source

Safety

WARNING

Hazard: Tilt drive commanded while you observe whether the motor rotates

Consequence: Crush / pinch injury to hands, clothing, or the couch if they are near the power cylinder, pillow block, or gantry frame when tilt moves

Avoidance: Clear hands, tools, and the couch from the tilt path before commanding tilt to run the check; keep a hand near the E-stop

CAUTION

Hazard: Stored mechanical energy in the gas spring and in a loaded power cylinder

Consequence: The gas spring or cylinder can move suddenly when disturbed, pinching fingers or striking the technician

Avoidance: Support the tilt frame/cover before disconnecting the gas spring or opening the power cylinder; never work under an unsupported tilt frame

NOTICE

Hazard: Commanding tilt with the transportation (shipping) jig still installed

Consequence: The jig blocks the mechanism and the drive can be damaged if it is run against it

Avoidance: Confirm the transportation jig is removed before testing or commanding tilt

Full step-by-step for Diagnose gantry tilt faults — tilt disabled or noisy during tilt operation

All 0 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry guides

View all Activion16 (TSX-031A) guides

Source: 2D201-112EN*D §1.2.1 Troubleshooting and repair of the tilt section (pp.35–36). Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.