Skip to main content
ProcedureGantrySetup & CalibrationEmergency

Release the gantry rotation brake to turn the rotation section by hand (emergency)

Use this when the Activion16 gantry rotation section must be turned by hand — for service work, or when the X-ray tube is not at its normal stop position and the powered recall (via the operating panel) is not available. There are two ways to reposition the rotation section: a powered recall to the home position (normal use, needs the console active) and a manual brake release (service/emergency u

Applies to
Activion16 (TSX-031A). Manual brake release is a service/emergency operation on the gantry rotation drive; no rotation-variant restriction is noted in the source for this task.
Steps
9 steps
Source
2D201-112EN*D §3.7.1 (p.254-255) — Stop position and methods for brake release

Tools & parts

ItemPart numberQty
None — uses the gantry's own breakers/switch (CP1, CP320, SW326, NFB2)n/a in source—

Safety

WARNING

Hazard: Attempting to turn the gantry rotation section by hand while the servo amplifier breaker CP320 is still ON

Consequence: (derived - not in source) The rotation motor still has power and can fight or drive the manual turning, risking injury or drive/brake damage

Avoidance: Always set CP320 to OFF (Step 2) before releasing the brake at SW326 — the source states Step (2) is mandatory to ensure safety

CAUTION

Hazard: Rotor-side electronics (HFG, DAS, etc.) left powered while the rotation section is opened up for manual turning

Consequence: (derived - not in source) Unnecessary exposure to live rotor electronics during service work

Avoidance: If power to the inside of the rotor is not needed for the task, also set breaker NFB2 (rear lower right of the gantry) to OFF (source: 'be sure to set breaker NFB2 ... to OFF')

NOTICE

Hazard: Turning the gantry rotation section by hand without first releasing the brake at SW326

Consequence: Brake malfunction / damage to the brake mechanism

Avoidance: Always complete the brake release procedure (CP1 ON, CP320 OFF, SW326 up) before turning the rotation section manually

Full step-by-step for Release the gantry rotation brake to turn the rotation section by hand (emergency)

All 9 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

7-day free trial · no credit card required

Related Gantry guides

View all Activion16 (TSX-031A) guides

Related error codes

Source: 2D201-112EN*D §3.7.1 (p.254-255) — Stop position and methods for brake release. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.