Pular para o conteúdo principal
ProcedureGantryReplacementReplace

Remove and refit the Activion16 gantry top cover

Remove the gantry top cover — one of the six lettered access covers in the gantry cover map (§1.3 overview, labeled [E]: Top cover; see figure, p.44) — to reach the maintenance sections underneath it, then refit the cover when the work is done. The cover is held by 4 M4 screws; removal is a straight lift, no other cover or cable disconnection is called out for this task.

Applies to
Activion16 (TSX-031A). No model- or revision-specific restriction is stated in source for this task.
Personnel
1 technician · Field service engineer (basic mechanical service)
Steps
3 steps
Source
2D201-112EN*D §1.3.6 (p.55) — Removing/Mounting the top cover

Tools & parts

ItemPart numberQty
Gantry top cover (resin panel)1
M4 screws, cover mounting4

Safety

DANGER

Hazard: Servicing performed with a gantry cover open

Consequence: Electric shock or injury from unexpected system operation

Avoidance: Turn OFF system power at the distribution board and wait 5 minutes or more before starting (general precaution G-D02; not restated per-cover in source §1.3.6, but applies whenever a gantry cover is open)

WARNING

Hazard: Cover mounting screws left loose after finishing work

Consequence: The cover can open, fall, or strike nearby objects — injury or equipment damage, especially with the gantry tilted

Avoidance: Confirm all 4 M4 screws are tightened firmly before returning the system to service (general precaution G-W11)

Full step-by-step for Remove and refit the Activion16 gantry top cover

All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry guides

View all Activion16 (TSX-031A) guides

Source: 2D201-112EN*D §1.3.6 (p.55) — Removing/Mounting the top cover. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.