Remove and refit the Activion16 gantry top cover
Remove the gantry top cover — one of the six lettered access covers in the gantry cover map (§1.3 overview, labeled [E]: Top cover; see figure, p.44) — to reach the maintenance sections underneath it, then refit the cover when the work is done. The cover is held by 4 M4 screws; removal is a straight lift, no other cover or cable disconnection is called out for this task.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Gantry top cover (resin panel) | — | 1 |
| M4 screws, cover mounting | — | 4 |
Safety
Hazard: Servicing performed with a gantry cover open
Consequence: Electric shock or injury from unexpected system operation
Avoidance: Turn OFF system power at the distribution board and wait 5 minutes or more before starting (general precaution G-D02; not restated per-cover in source §1.3.6, but applies whenever a gantry cover is open)
Hazard: Cover mounting screws left loose after finishing work
Consequence: The cover can open, fall, or strike nearby objects — injury or equipment damage, especially with the gantry tilted
Avoidance: Confirm all 4 M4 screws are tightened firmly before returning the system to service (general precaution G-W11)
Full step-by-step for Remove and refit the Activion16 gantry top cover
All 3 steps are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry guides
Source: 2D201-112EN*D §1.3.6 (p.55) — Removing/Mounting the top cover. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.