TUBE-FIL-LIMIT — Exposure technique exceeds the X-ray tube filament operating limit
Philips CT error code, reported by the Generator (GEST).
What TUBE-FIL-LIMIT means
On DuraDiagnost F30 DR CT systems, status code TUBE-FIL-LIMIT indicates: Exposure technique exceeds the X-ray tube filament operating limit. The code originates in the X-ray Generator (GEST) section of the Generator (GEST). Severity level warning (Warning).
Reported action: Increase the kV, reduce the mA, or switch to the large focal spot, per the OEM message..
From the Generator "Messages to Operator and User" table in the Philips DuraDiagnost F30 DR Technical Manual (DMR293226 REV 04). Shown on the GEST PC Tools screen / GEST Generator Report when the requested exposure parameters go beyond the maximum operating limit of the selected X-ray tube filament.