Alarm — control panel faultOperator Control Panel (Left/Right)

PERRControl panel error PErr — the Control Panel CAN cable is mistakenly plugged into the Patient Table CAN port on the Gantry Interface Board

Philips CT error code, reported by the Operator Control Panel.

What PERR means

On MX 16-slice CT CT systems, status code PERR indicates: Control panel error PErr — the Control Panel CAN cable is mistakenly plugged into the Patient Table CAN port on the Gantry Interface Board. The code originates in the Operator Control Panel (Left/Right) section of the Operator Control Panel. Severity level alarm (Alarm — control panel fault).

Reported action: "PErr" on the Control Panel display indicates the Control Panel CAN cable was mistakenly plugged into the Patient Table CAN port on the Gantry Interface Board. Verify and correct the CAN cable connections at the Gantry Interface Board so the Control Panel connector is in the correct CAN port..

Documented in the Philips MX 16-slice CT Service Troubleshooting Guide (455019303481 Rev. G), Table 4 "Control Panel Error Code", as a code shown on the (Left/Right) Control Panel display. Not confirmed by this document for other Philips CT systems.

Full troubleshooting for PERR