Pular para o conteúdo principal
Alarm — control panel faultOperator Control Panel (Left/Right)

CERRControl panel error CErr — the panel does not receive the signal from the UCOM

Philips CT error code, reported by the Operator Control Panel.

What CERR means

On MX 16-slice CT CT systems, status code CERR indicates: Control panel error CErr — the panel does not receive the signal from the UCOM. The code originates in the Operator Control Panel (Left/Right) section of the Operator Control Panel. Severity level alarm (Alarm — control panel fault).

Reported action: "CErr" on the Control Panel display indicates the panel is not receiving a signal from the UCOM. Check the CAN/communication cabling and connectors between the Control Panel and the UCOM board, and confirm the UCOM board is powered and operating normally..

Documented in the Philips MX 16-slice CT Service Troubleshooting Guide (455019303481 Rev. G), Table 4 "Control Panel Error Code", as a code shown on the (Left/Right) Control Panel display. Not confirmed by this document for other Philips CT systems.

Full troubleshooting for CERR

The complete diagnosis, cause tree, and step-by-step remedy for the Operator Control Panel (Left/Right) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Operator Control Panel (Left/Right) codes

Discuss CERR with other field engineers in the FSELIB forum

Source: Philips MX 16-slice CT Service Troubleshooting Guide — 455019303481 Revision G (Preliminary, 3 February 2016). Reference information re-authored by FSELIB — not a copy of the OEM manual.