CERR — Control panel error CErr — the panel does not receive the signal from the UCOM
Philips CT error code, reported by the Operator Control Panel.
What CERR means
On MX 16-slice CT CT systems, status code CERR indicates: Control panel error CErr — the panel does not receive the signal from the UCOM. The code originates in the Operator Control Panel (Left/Right) section of the Operator Control Panel. Severity level alarm (Alarm — control panel fault).
Reported action: "CErr" on the Control Panel display indicates the panel is not receiving a signal from the UCOM. Check the CAN/communication cabling and connectors between the Control Panel and the UCOM board, and confirm the UCOM board is powered and operating normally..
Documented in the Philips MX 16-slice CT Service Troubleshooting Guide (455019303481 Rev. G), Table 4 "Control Panel Error Code", as a code shown on the (Left/Right) Control Panel display. Not confirmed by this document for other Philips CT systems.