ER29 — Patient table error (grouped code)
Philips CT error code, reported by the Patient Table / Patient Support.
What ER29 means
On Access Series CT (6/16/Dual) CT systems, status code ER29 indicates: Patient table error (grouped code). The code originates in the Patient Table / Patient Support section of the Patient Table / Patient Support. Severity level alarm (Alarm — stator control board error condition).
Reported action: Check the System Error Code in the Error Log Viewer/system log (one of 0x5002-0x5018 listed) to identify the specific patient table fault, then inspect the corresponding Patient Table Motion Control Board LED indicators (Section 1.10/1.11) and servo error codes (Table 1-12)..
Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family; maps to System Error Codes 0x5002, 0x5003, 0x5006, 0x5007, 0x5009, 0x5011, 0x5012, 0x5014, 0x5015, 0x5016, 0x5017, 0x5018. Not confirmed for other Philips CT systems.
Full troubleshooting for ER29
The complete diagnosis, cause tree, and step-by-step remedy for the Patient Table / Patient Support subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Patient Table / Patient Support codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.