Alarm — servo drive faultPatient Table / Patient Support

ALE01Patient table servo drive over-current

Philips CT error code, reported by the Patient Table Servo Drive.

What ALE01 means

On Access Series CT (6/16/Dual) CT systems, status code ALE01 indicates: Patient table servo drive over-current. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).

Reported action: Check the patient table servo motor and drive wiring for a short circuit or mechanical bind causing excess current draw; inspect the servo drive fuses/overcurrent protection before replacing the drive..

Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ALE01

The complete diagnosis, cause tree, and step-by-step remedy for the Patient Table / Patient Support subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Patient Table / Patient Support codes

Discuss ALE01 with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.