ALE06 — Patient table servo overload
Philips CT error code, reported by the Patient Table Servo Drive.
What ALE06 means
On Access Series CT (6/16/Dual) CT systems, status code ALE06 indicates: Patient table servo overload. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).
Reported action: Check for excessive mechanical load or binding on the patient table (patient weight limits, obstructions, worn mechanics) before replacing the servo drive..
Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.
Full troubleshooting for ALE06
The complete diagnosis, cause tree, and step-by-step remedy for the Patient Table / Patient Support subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Patient Table / Patient Support codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.