Alarm — stator control board error conditionSlip Ring Communication

ER23Stator Control Board failed to send 800us status frame

Philips CT error code, reported by the Slip Ring Communication.

What ER23 means

On Access Series CT (6/16/Dual) CT systems, status code ER23 indicates: Stator Control Board failed to send 800us status frame. The code originates in the Slip Ring Communication section of the Slip Ring Communication. Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5714 in the system log. Check the slip ring link and Stator Control Board before replacing hardware..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Note: this table also defines a distinct servo-drive code 'Err23' in Table 1-12 with a different meaning. Not confirmed for other Philips CT systems.

Full troubleshooting for ER23